<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>45826</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: There is no plate detected on the fetcher where one was expected.\n  Plates in cassette &lt;Cassette Number&gt; may be pinched by the side rails.
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Magnus_VLF</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The plate sensor on the fetcher is not reading a plate but one was expected.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Check that plates are not pinched by cassette side rails.  There should be a 1 mm
gap between the plate stack and the side rails and end stops.  The plates should be
free to move.
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
The picker failed to lift the plate (possibly due to cassette interference).
</Cause>
<Cause>
The picker dropped the plate.
</Cause>
<Cause>
Faulty sensor.
</Cause>
<Parameter>
Cassette Number -
This is the cassette number.
</Parameter>
<LastModified>
27-Apr-2004
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:47 2008 PST
</Updated>
<Links>
</Links>
</Error>